Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Soltec A11 (1 year) CAL VER /5500 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 136 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- N K P F V 1.003 ASK+ R X 1.004 HEAD INITIAL SETUP 1.005 DISP On the Mainframe, set the following... 1.005 DISP 1.005 DISP TRIGGER: MANU. TRIG. 1.005 DISP Y SIZE: 200mm 1.005 DISP T SIZE: X1 1.006 SET TIME/DIV: 1 sec. 1.007 DISP EXTERNAL CONTROL settings on the Mainframe... 1.007 DISP MODE: REAL 1.007 DISP TRIGGER: SHOT 1.008 DISP On the A11, set the following... 1.008 DISP 1.008 DISP LINE WIDTH: MEDIUM 1.008 DISP TRIGGER: OFF 1.008 DISP RANGE/DIV: 5 mV 1.008 DISP POSITION %: 0 1.008 DISP FILTER: 0FF 1.009 HEAD {POSITION} 1.010 DISP Connect the 5500 AUX OUTPUT to the A11 CHANNEL INPUT. 1.011 5500 0V S 2W 1.012 HEAD POSITION: {0%} 1.013 DISP Set A11 POSITION %: 0 1.014 EVAL Does the {trace move 0% towards the positive}? 2.001 HEAD POSITION: {10%} 2.002 5500 0V S 2W 2.003 DISP Set A11 POSITION % to 10 2.004 EVAL Does the {trace move 10% towards the positive}? 3.001 HEAD POSITION: {20%} 3.002 5500 0V S 2W 3.003 DISP Set A11 POSITION % to 20 3.004 EVAL Does the {trace move 20% towards the positive}? 4.001 HEAD POSITION: {30%} 4.002 5500 0V S 2W 4.003 DISP Set A11 POSITION % to 30 4.004 EVAL Does the {trace move 30% towards the positive}? 5.001 HEAD POSITION: {40%} 5.002 5500 0V S 2W 5.003 DISP Set A11 POSITION % to 40 5.004 EVAL Does the {trace move 40% towards the positive}? 6.001 HEAD POSITION: {50%} 6.002 5500 0V S 2W 6.003 DISP Set A11 POSITION % to 50 6.004 EVAL Does the {trace move 50% towards the positive}? 7.001 HEAD POSITION: {60%} 7.002 5500 0V S 2W 7.003 DISP Set A11 POSITION % to 60 7.004 EVAL Does the {trace move 60% towards the positive}? 8.001 HEAD POSITION: {70%} 8.002 5500 0V S 2W 8.003 DISP Set A11 POSITION % to 70 8.004 EVAL Does the {trace move 70% towards the positive}? 9.001 HEAD POSITION: {80%} 9.002 5500 0V S 2W 9.003 DISP Set A11 POSITION % to 80 9.004 EVAL Does the {trace move 80% towards the positive}? 10.001 HEAD POSITION: {90%} 10.002 5500 0V S 2W 10.003 DISP Set A11 POSITION % to 90 10.004 EVAL Does the {trace move 90% towards the positive}? 11.001 HEAD POSITION: {100%} 11.002 5500 0V S 2W 11.003 DISP Set A11 POSITION % to 100 11.004 EVAL Does the {trace move 100% towards the positive}? 12.001 DISP Set A11 POSITION % to 50 12.002 HEAD {FULL SCALE} 12.003 MESS Slew for Full Scale (10 div.) Deflection. 12.004 HEAD FULL SCALE: {5 mV/div Range} 12.005 5500 5 50.00mV 1% 0.195U 10H SI 2W 13.001 HEAD FULL SCALE: {10 mV/div Range} 13.002 5500 10 100.00mV 1% 0.39U 10H SI 2W 14.001 HEAD FULL SCALE: {20 mV/div Range} 14.002 5500 20 200.00mV 1% 0.781U 10H SI 2W 15.001 HEAD FULL SCALE: {50 mV/div Range} 15.002 5500 50 500.00mV 1% 1.95U 10H SI 2W 16.001 HEAD FULL SCALE: {.1 V/div Range} 16.002 5500 0.1 1.00V 1% 0.0039U 10H SI 2W 17.001 HEAD FULL SCALE: {.2 V/div Range} 17.002 5500 0.2 2.00V 1% 0.0078U 10H SI 2W 18.001 HEAD FULL SCALE: {.5 V/div Range} 18.002 5500 0.5 5.00V 1% 0.019U 10H SI 2W 19.001 HEAD FULL SCALE: {1 V/div Range} 19.002 5500 1 10.00V 1% 0.039U 10H SI 2W 20.001 HEAD FULL SCALE: {2 V/div Range} 20.002 5500 2 20.00V 1% 0.078U 10H SI 2W 21.001 HEAD FULL SCALE: {5 V/div Range} 21.002 MESS Slew for 50% Scale (5 div.) Deflection. 21.003 5500 5 25.00V 1% 0.097U 10H SI 2W 22.001 HEAD {LINE WIDTH} 22.002 HEAD LINE WIDTH: {Off} 22.003 MESS Set the LINE WIDTH control to OFF. 22.004 EVAL [N]Is {there a line recording}? 23.001 HEAD LINE WIDTH: {Light} 23.002 MESS Set the LINE WIDTH control to LIGHT. 23.003 EVAL Is the {line recording Light}? 24.001 HEAD LINE WIDTH: {Medium} 24.002 MESS Set the LINE WIDTH control to MEDIUM. 24.003 EVAL Is the {line recording Medium}? 25.001 HEAD LINE WIDTH: {Bold} 25.002 MESS Set the LINE WIDTH control to BOLD. 25.003 EVAL Is the {line recording Bold}? 26.001 MESS 26.002 DISP Press STOP on the Mainframe to stop the chart paper. 26.003 END